DAD 141.1 Technical Manual, Rev. 2.2 - February 2014
Page 15 of 54
7.3. Menu 1 System Zero
Remark: Activate a new calibration with 1x Power OFF/ON !
1.
ZERO setup (Menu 1.1 to 1.4) TAC protected see chapter 4.12
1.1
Automatic Zero Tracking - Enable / Disable (command ZT)
-
If enabled: Zero setting via button Zero or by external signal is
possible (Menu 6.1.1).
-
Settings: 0 … 255d
-
Disable @ 00000
Enable @ 00001 or higher (max 00255)
-
Setting 00001 sets a zero track range of ±0.5d
-
Setting 00002 up to 00255 sets a zero track range of
±1d up to ±127.5d, independent of decimal point setting
1.2
Calibrate system ZERO - gravimetric by weight / load (command CZ)
-
Display shows the actual input signal in mV/V.
Press ENTER button to store ZERO.
Remark: Scale should/must be unloaded.
1.3
Calibrate system ZERO - electronic by mV/V value (command AZ)
-
Use the UP/DOWN & MOVE RIGHT keys to setup the mV/V value
at which the device should read ZERO
1.4
System ZERO & TARE function
1.4.1
Store TARE value non volatile: ON / OFF (command TN)
-
ON: store non-volatile @ power OFF
-
OFF: delete @ power OFF
1.4.2
Store ZERO value non volatile: ON / OFF (command ZN)
-
ON: store non-volatile @ power OFF
-
OFF: delete @ power OFF
1.4.3
Initial ZERO @ power ON: ON / OFF (command ZI)
-
ON: proceed initial Zero @ power ON
-
Range is ±10% of Max
1.4.4
ZERO range (increments) (command ZR)
-
Defines the zero setting range in divisions. If this parameter is set
to zero, a zero setting range of ±2% of Max is used, independent
of decimal point setting.